64
Keysight U3042AM20 User’s and Service Guide
U3042AM20
Cal Kit Operational Check
-
Figure 45
Calibration, Cal Set Viewer
10. From the Cal Sets drop-down menu, select 999.1 and select Enable.
Select the Reflection Tracking(x,x), where x,x is the port being tested, term
in the center drop-down menu and ensure that the Enable and Error Terms
are selected.
11. Compare the Reflection Tracking (1,1) trace to the appropriate limits in
. Use the Marker function to measure trace values.
The trace should be similar to
.
12. Repeat step 10 above. Compare the Reflection Tracking (1,1) trace to the
appropriate limits in
. Use the Marker function to
measure trace values. The trace should be similar to
for Cal Sets “999.1” thru “999.24” (24-Port).
Response from 10 MHz to 500 MHz is normal due to the PNA-X couplers in
comparison to the Test Set bridges. The bridges have more gain in the
coupled RF path.