Keysight U3042AM20 User’s and Service Guide
U3042AM20
Troubleshooting the Test Set
-
Equipment Setup
1. Turn on the test set and the analyzer.
2. Preset the analyzer and set it to Standalone Mode. Select Utility > System
>
Configure > Multiport Capability. In the dialog box select Restart as a
standalone PNA > OK.
3. Confirm the frequency range is set to 10 MHz to 26.5 GHz.
4. Connect the RF cables to the analyzer’s Ports 1 and 2. Connect the cables
together using a 3.5 mm adapter.
5. Configure the analyzer to measure S21 and normalize the response trace.
6. Set the analyzer to Interface Control Mode: Select Channel > Hardware
Setup > More > Interface Control… and click Enable Interface Control box.
The <addrs>.<data> entries noted in the following Test Instructions table will
be used to configure the RF switches for this testing. After making your entry
select <OK> to execute the command, to return back for further entries,
select Interface Control on the PNA display.
Table 20
RF Signal Path Insertion Loss (S21)
RF Path Description
Signal
Insertion Loss (typical)
Source IN to Ports 5-24
Source
RCVR OUT to Ports 5-24
Receiver
Source IN to CPLR THRU
Source
RCVR OUT to CPLR ARM
Receiver