Keysight U3042AM20 User’s and Service Guide
43
U3042AM20
Controlling the Test Set with the Analyzer
Using Interface Control Mode
An Instrument Preset will reset all of the fields to their default settings.
If an error is encountered when sending Interface Control data, an error
message is displayed on the analyzer’s screen and the Channel Trigger State
is set to Hold. You must fix the condition that caused the error, then change
the Channel Trigger State to its original setting.
Figure 34
Interface Control
Enable Interface Control: (1)
Enables and disables ALL Interface Control communication. When cleared
(default setting) Interface Control is disabled and NO data is sent. To send
data, the individual interfaces must also be enabled.
Test Set I/O Control (addr.data): (2)
Provides control of the test set I/O Interface on the rear panel of the
analyzer. Used to control your test set.
Only positive integers are allowed to select switch positions or states of DUT
control interface lines. Refer to
“Address and Data Values” on page 47
.
Address and data integers must be separated by a period. Each set of entries
should be separated by a new line, or carriage return. The front panel Enter
key can be used to insert a new line into the field. The quantity of test set I/O
entries that can be entered is limited by the available memory of the
analyzer.
Address and Data example: addr.data (3)
0.0
16.1
32.2
(1)
(2)
(3)
(4)
(5)
(7)
(8)
(8)
(6)