Keysight U3042AM20 User’s and Service Guide
47
U3042AM20
Address and Data Values
Address and Data Values
Setting the Test Port Paths with Address and Data
Refer to
,
information to set the internal switch paths of the test set.
The address is the first value in the test set I/O control or GPIB data command.
The second value controls the source and receiver paths of the ports. To select
a test set port configuration both switches must be set to complete the source
and receiver paths. To do this you must add the data values together to
determine the data command value for each port, which is the second number
in the Test Set I/O or GPIB command. Refer to
and in
“SCPI Control Mode” on page 45
Example 1:
Refer to
If the ports have different addresses, two separate address data commands
must be used.
Port 5 is the Source and Port 6 is the Receiver.
Source Port 5 = address 0, data 1 and Receiver Port 6 = address 16, data 16.
Two separate commands must to be sent, you may use the same dialog box.
The entry will be 0.1 in one command line, and 16.16 in the second line.
Example 2:
If the ports have the same address, only one command is needed.
Port 1 is the Source and Ports 5 is the Receiver.
Source Port 1 = address 0, data 0 and Receiver Port 5 = address 0, data 16. The
data values are added together, the entry will be 0.16.
Example 3:
(Option 129) Refer to
Port 9 is the Source and Ports 5 is the Receiver.
Noise Source Port 5 = address 0, data 9, and Receiver Port 9 = address 0, data
16.
Two separate commands must be sent, you may use the same dialog box. The
entry will be 0.9 in one command line, and 16.16 in the second line.