38
Keysight U3042AM20 User’s and Service Guide
U3042AM20
Controlling the Test Set with the Analyzer
-
If the analyzer has Option 029 or H29 (noise figure capability), verify that the
path configuration is set to Default. If it is not in Default configuration, ports
5 and 9 will not operate correctly.
Select
Trace/Chan > Channel > Hardware Setup > Path Config...
-OR-
(Instrument > Setup > Internal Hardware > Path Config...)
and in the drop-
down menu select Default > OK. Save this configuration as a User Preset by
selecting Save > User Preset > Save current state as User Preset. Do not use
the factory Preset (User Preset Off), the analyzer will return to Option 029
path configuration.
Figure 30
RF Path Configuration