Keysight U3042AM20 User’s and Service Guide
U3042AM20
Controlling the Test Set with the Analyzer
-
Controlling the Test Set with the Analyzer
This section will describe how to operate the test set with the analyzer.
The internal firmware of the analyzer has not been modified for this test set
option. Power levels may differ from those indicated on the network analyzer
when the test set is connected.
The test set is considered a “slave” instrument. A PNA, PNA-X or PNA-L must
be used to control the test set. There are three methods to control the test set.
Multiport mode is recommended due to calibration and ease of use.
—
“Multiport Mode (Option 551)” on page 32
.
—
“Interface Control Mode” on page 41
.
—
“SCPI Control Mode” on page 45
Typeface Key Conventions
The following key conventions are used throughout this document.
—
[HARDKEYS]
are labeled front panel keys
—
SOFTKEYS
are indicated on the instrument display
—
(Italicized in parenthesis)
are menu paths for the 'B' Model analyzer
Sweep Setup for Multiport and Standalone Modes
When the Test Set is connected to the network analyzer, it is recommended
that the analyzer’s Sweep Setup be configured to Stepped Sweep before
calibrating. This is slower than the Analog Sweep, but is more accurate due to
the extra electrical length of the Test Set and test port cables. Stepped Sweep
is available on all models.
1.
On the analyzer select STIMULUS > Sweep > Sweep Setup.
2.
Select Stepped Sweep (sweep moves in discrete steps).
3.
Set the Dwell Time to 5 μs > OK.