Keysight U3042AM20 User’s and Service Guide
41
U3042AM20
Controlling the Test Set with the Analyzer
Interface Control Mode
Interface Control mode will not function properly when using multiport mode.
Multiport mode will reset the switch path commands of the interface control.
It is recommended that the network analyzer be restarted in stand-alone
mode if the interface control is being used.
Overview of the Interface Control
The Interface Control feature allows you to send data and remote commands to
control external equipment using the GPIB, Handler I/O, Test Set I/O, Dwell
After Command and Aux I/O without needing to create a remote program.
Refer to PNA Help menu, “Rear Panel Tour.”
— A unique set of control data can be sent for each channel. In addition, a
unique set of control data can be sent before the channel sweep starts and
after the sweep ends.
— Interface Control settings can be saved and recalled from the Interface
Control dialog box or with Instrument State Save and Recall.
— Interface Control settings can be copied to other channels using Copy
Channels.
— Control data can only be WRITTEN to the interfaces, NOT READ from the
interfaces.
— Control data is sent in the following order and this order cannot be changed
(Refer to the PNA Help menu):
1.
GPIB Commands - BEFORE
2.
Handler I/O Control
3.
Test Set I/O Control (addr.data)
4.
Dwell After Command
5.
Aux I/O Output Voltage