Installation Note N9010-90054
3
Option HL4, 40 MHz Analysis Bandwidth (B40) + Microwave Preselector Bypass (MPB) Upgrade Kit
Option HL4, 40 MHz Analysis Bandwidth (B40) + Microwave
Preselector Bypass (MPB) Upgrade Kit
Software and test equipment is required for making adjustments and for performance verification
testing.
Information on how to obtain this software can be found at:
www.keysight.com/find/calibrationsoftware
While Keysight does recommend that a full calibration be performed after the installation of this
upgrade, the end user must ultimately determine whether they want this service or not. If a full
calibration is required, arrangements regarding the level of calibration must be made between the
end user and the calibration provider.
Products Affected:
N9010A, EXA Signal Analyzer
N9020A, MXA Signal Analyzer
Serial Numbers:
N9010A,
<
MY/SG/US56480000
N9020A,
<
MY/SG/US52330000
Options: B40,
MPB
To Be Performed By:
(X) Keysight Service Center
(X) Personnel Qualified by Keysight
a
( ) Customer
a. To ensure that these newly installed options are functioning properly, the procedure that follows
includes the requirement of performing certain adjustments and performance verification tests.
However, the completion of these tests does not guarantee that the instrument meets all advertised
specifications.
Estimated Installation Time:
Estimated Adjustment Time:
Estimated Verification Time:
3.0 Hours
5.5 Hours
3.5 Hours