Installation Note N9010-90054
59
Installation Procedure
Loading New Hardware Drivers
On the first boot-up after installing the new hardware, the analyzer will detect the new hardware
and will need to load drivers for the new hardware. The analyzer will boot-up as Instrument user,
but loading the new drivers requires administrator privileges.
1.
Turn on the analyzer and wait for the XSA application to launch.
2.
During the start up of the XSA application, it will detect the new hardware and need to load
new drivers. You will be prompted for an administrator login and password. The default
administrator login is “administrator” and the default password is “agilent4u”.
3.
You will receive two prompts as described above. The first prompt is for the “Agilent PXA DIF”
and the other is for the “Agilent Receiver Module”. Both of these relate to the new A3 40 MHz
Digital IF Assembly.
4.
After the hardware drivers have been installed, the XSA application will continue to start up.
Continue to Verify Hardware Installation.
Update FPGA Code
On the first boot-up after installing the new hardware, the analyzer will detect the new A3 40 MHz
BW Digital IF and A15 Front End Controller. However, the FPGA code in the new assemblies will not
necessarily be synchronized with the instrument software installed earlier. The following procedure
ensures that the FPGA code in the new assemblies is updated to the correct revision.
1.
Turn on the analyzer and wait for the XSA application to launch.
2.
Close the instrument application software by pressing
File
,
Exit
,
Enter
using the front panel
keys.
3.
Connect a USB mouse to one of the instrument front panel USB ports.
4.
Using the mouse, click
Start
,
Log Off
,
Log Off
.
5.
Relogin as “administrator”. The default password is “agilent4u”. Click
OK
.
6.
Using the mouse, double-click on the Windows Explorer link on the desktop and navigate to
the following folder:
C:\Program Files\Keysight\SignalAnalysis\Physics
7.
In this folder find and execute the file named:
FPGA_Prog.bat