44
Installation Note N9010-90054
Installation Procedure
7.
Refer to
. To pair the AIF and DIF together again, set the DIF
(2)
in position over the
AIF
(1)
. Attach the AIF to the DIF with three screws
(3)
(0515-0372) included in the kit.
Torque to 9 inch-pounds, starting with the middle screw.
Figure 31
AIF/DIF Pairing
8.
Connect the ribbon cable W2 to J820 on the AIF
(1)
.
9.
Slide the AIF/DIF assembly into the slot at the rear of the instrument and push on the
assembly to mate the connectors to the motherboard assembly.