14
Installation Note N9010-90054
Installation Procedure
Analyzer Disassembly
Standard Instrument (Benchtop Configuration)
1.
Disconnect the instrument from ac power.
2.
Refer to
. Using the T-20 driver, remove the four screws (two on each side)
(1)
that
attach the handle strap
(2)
on each side of the instrument.
3.
Using the T-20 driver, remove the four screws (including washers)
(3)
that hold the rear feet
(4)
in place.
4.
Pull the instrument cover
(5)
off towards the rear of the instrument.
If the instrument is placed on its face during any of the following procedures, be sure to use a soft
surface or soft cloth to avoid damage to the front panel, keys, or input connector.
If the analyzer has Option PRC, Portable Configuration, refer to the
to remove the outer case.
Make sure any adapters on the front panel are removed.