Installation Note N9010-90054
9
Installation Procedure
6.
Refer to the data in
above. If the analyzer has N90x0A-503, N9010A-532, or
N9010A-544
do not proceed
with the installation of this kit. This kit can only be installed in
analyzers with frequency range options 507, 508, 513, and 526.
7.
Refer to the data in
above. If the analyzer is an N9010A and the serial number is
≥
MY/SG/US56480000
do not proceed
with the installation of this kit, This kit is only for older
N9010A and N9020A analyzers.
8.
Refer to the data in
above. If the analyzer is an N9020A and the serial number is
≥
MY/SG/US52330000
do not proceed
with the installation of this kit, This kit is only for older
N9010A and N9020A analyzers.
9.
Refer to the data in
above. If the Instrument S/W Revision is earlier than A.10.05, you
must upgrade the instrument software. The hardware and licenses being added with this
upgrade are only supported by software revisions A.10.05 and later.
Even if the Instrument S/W Revision is A.10.05 or later, we recommend upgrading to the latest
software revision to ensure that you get the benefit of any defect fixes that have been made
since A.10.05.
The latest instrument software can be downloaded from the following web site:
http://www.keysight.com/find/xseries_software
10.
Refer to the data in
above. If the Front End has a Hw Id of 70, it already has the 150 MHz
wide 1st IF bandpass filter and you do not need to perform the Replace A13A1 Bandpass Filter
procedure.
11.
above. If the analyzer has N90x0A-CR3 and/or
N90x0A-CRP installed, verify that it also has a Front End Controller with Hw Id of 41 or 75. If
the Front End Controller has a Hw ID of 41 or 75, you do not need to perform the Replace Front
End Controller procedure.