30
Installation Note N9010-90054
Installation Procedure
3.
. Install the Enhanced Front End Control assembly included in this kit into
slot 6 in the chassis securing with the ejector.
4.
Reattach the ribbon cables W16, W17, and W18 to the Front End Control assembly
(1)
.
5.
If the analyzer does not have Option 503, replace the existing W22 wire harness with the new
W22 (
) included in the kit. Connect W22 between J300 on the Front End Control
and the A12 YTF Preselector.
Figure 17
Enhanced Front End Control Assembly Installation