Installation Note N9010-90054
13
Installation Procedure
Alternate Installation Procedure
The License File can be manually installed over USB or LAN by placing the license file in the
following analyzer folder: C:\Program Files\Agilent\licensing.
Verify the License Installation
1.
Before the licenses will be recognized, the XSA application must be restarted. Press
File
,
Exit
.
An Exit Analyzer dialog box will appear; press Enter to confirm the exit.
2.
Double-click on the LaunchXSA icon on the Windows desktop. Wait for the XSA application to
finish starting (the analyzer should be sweeping).
3.
Press
System
,
Show
,
System
on the analyzer to display a list of all displayed options. You
should see the following options listed (“x” is 1 for an EXA and 2 for an MXA):
—
N90x0A-B40 Analysis bandwidth, 40 MHz
—
N90x0A-B25 Analysis bandwidth, 25 MHz
—
N90x0A-DP2 Digital Processor, 2 GB Deep Capture Memory
—
N90x0A-MPB Microwave
Preselector
Bypass
—
N90x0A-RTL Real Time Link
—
N90x0A-S40 Baseband
I/Q
Analysis bandwidth, 40 MHz (MXA only)
—
N9060A-3FP Amplitude Correction and Limit Lines
—
N9060A-4FP TOI and Harmonics
—
N9060A-5FP ACP 18 Carrier Measurement
—
N9060A-6FP TV Trigger and IQ Analyzer Enhancement
4.
Press
System
,
Show
,
Errors
,
Clear Error Queue
. This will remove out-of-date errors.
5.
Press
System
,
Alignments
,
Align Now
,
All
.
6.
Several alignments will be performed related to optional functionality you just installed.
7.
Watch for any errors during the alignments.
8.
Press
System
,
Show
,
Errors
to view the instrument error log.
9.
Verify that there were no errors during alignment.