Installation Procedure
Reassembly
1.
This kit contains two RF Brackets. One bracket is for older analyzers and the other is for newer
analyzers. Select the appropriate bracket based upon the model and serial number of your
analyzer.
2.
Refer to
. Using the appropriate RF Bracket included in the kit, position the RF
bracket
(1)
onto the chassis and replace seven of the eight screws
(2)
(the previous RF
Bracket secured to the Low Band Switch Bracket using two screws; the new RF Bracket only
requires one screw). Torque to 9 inch-pounds.
Figure 40
RF Bracket Replacement
Model Number
Serial Number
RF Bracket Part Number
N9010A
< MY/SG/US52180000
N9010A
> MY/SG/US52180000
N9020A