Installation Note N9010-90054
61
Installation Procedure
Verify Hardware Installation
1.
Verify the spectrum analyzer application loads and sweeps as expected.
2.
Press
System
,
Show
,
Hardware
on the analyzer and verify that the DIF, EAIF, and EFEC boards
identify themselves as:
—
Analog IF (Hw Id = 57)
—
Digital IF (Hw Id = 72)
—
Front End Controller (Hw Id = 41 or 75)
Verify Optional Functionality
1.
Press
Mode
,
IQ Analyzer (Basic)
.
2.
Press
Mode Setup
.
3.
Press
IF Path
.
4.
Verify that softkeys appear labeled “
25 MHz
” and “
40 MHz
”.
5.
Press
AMPTD Y Scale
,
More 1 of 2
,
uW Path Control
.
6.
Verify that softkeys appear labeled “
Standard Path
” and “
uW Preselector Bypass
”.
7.
Press
uW Preselector Bypass
. If the path was previously set to
Standard Path
, you should hear
a click.
8.
Press
Standard Path
. If the path was previously set to
uW Preselector Bypass
path, you should
hear a click.
Perform Preselector Characterization
1.
Press
System
,
Alignments
,
More
,
Advanced
.
2.
Press
Characterize Preselector
,
Enter
. The characterization will take several minutes.
3.
Wait until the analyzer resumes sweeping.