Keysight Infiniium V-Series Oscilloscopes
Service Guide
3 Testing Performance
Vertical Performance Verification
Analog Bandwidth—Maximum Frequency Test
Time Scale Accuracy (TSA) Test
Full performance verification for V-Series oscilloscopes consists of three main
procedures:
1
Performing the internal oscilloscope self tests to ensure the measurement
system is functioning properly. To perform the self tests, Click
Utilities > Self
Test...
. Then select
Scope SelfTest
from the
Available Self Test
drop-down list box,
click
Start
, and follow the instructions on the screen. If any of the self tests fail,
ensure that the failure is diagnosed and repaired before calibrating and testing
performance.
2
Calibrating the oscilloscope, as described in chapter 2.
3
Testing the oscilloscope to ensure that it is performing to specification. This
chapter describes the performance test procedures.