Testing Performance
3
Infiniium V-Series Oscilloscopes Service Guide
37
Analog Bandwidth—Maximum Frequency Test
CAUTION
Ensure that the input voltage to the oscilloscope never exceeds
±5 V
.
NOTE
The oscilloscope under test must be warmed up (with the oscilloscope application running) for
at least 30 minutes prior to the start of any performance test.
Specifications
Analog Bandwidth (-3 dB)
MSO/DSO/DSAV084A 8.0 GHz
MSO/DSO/DSAV134A 13.0 GHz
MSO/DSO/DSAV164A 16.0 GHz
MSO/DSO/DSAV204A 20.0 GHz
MSO/DSO/DSAV254A 25.0 GHz
MSO/DSO/DSAV334A 32.0 GHz
Equipment Required
Description
Critical specifications
Recommended model/part #
Microwave
CW Generator
Maximum Frequency
≥
32 GHz
Power range: -20 dBm to +16 dBm into 50
Ω
Output resistance = 50
Ω
E8257D with Opt 540
Power Splitter 2 Resistor Power Splitter
Max Frequency
≥
32 GHz
11667C
Power Meter
Keysight E-series with power sensor
compatibility
E4418B or E4419B
Power Sensor
Maximum Frequency
≥
32 GHz
Power range: -24 dBm to +16 dBm
8487A, or E4413B with 8485A-033
Microwave
Cable
50
Ω
Characteristic Impedance
2.4 mm (m) to 2.4 mm (m) connectors
Max Frequency
≥
32 GHz
N5180-60204
Adapter
2.4 mm (m) to 3.5 mm (m) (qty. 1)
11901A
Adapter
2.4 mm (m) to 3.5 mm (f) (qty. 2)
11901C
Connector
Saver
3.5 mm (f)-(f), shipped with oscilloscope
54919-68717 (for models
≥
20 GHz)
54916-68716 (for models
≤
16 GHz)
Summary of Contents for Infiniium MSOV 084A
Page 1: ...Keysight Infiniium V Series Oscilloscopes Service Guide...
Page 6: ...6 Infiniium V Series Oscilloscopes Service Guide...
Page 14: ...14 Infiniium V Series Oscilloscopes Service Guide 1 General Information...
Page 22: ...22 Infiniium V Series Oscilloscopes Service Guide 2 Calibration...
Page 52: ...52 Infiniium V Series Oscilloscopes Service Guide 3 Testing Performance...
Page 132: ...132 Infiniium V Series Oscilloscopes Service Guide 5 Replacing Assemblies...
Page 162: ...162 Infiniium V Series Oscilloscopes Service Guide 7 Theory of Operation...
Page 163: ......