126
Infiniium V-Series Oscilloscopes Service Guide
5
Replacing Assemblies
Setting the calibration factors after replacing an acquisition board
After replacing an acquisition board you must perform a service calibration,
including the interleave response correction, and then perform a self test
procedure. Refer to Chapter 2, “Calibration”, for details.
NOTE
The oscilloscope under test must be warmed up (with the oscilloscope application running) for
at least 30 minutes prior to the start of any calibration or test.