Testing Performance
3
Infiniium V-Series Oscilloscopes Service Guide
39
6
Click
Control > Factory Default
, then configure the oscilloscope as follows:
a
Ensure the channel under test is displayed and all other channels are turned
off.
b
Set the vertical scale of the channel under test to 10 mV/div.
c
Set the horizontal scale to 16 ns/div (to display 8 cycles of a 50 MHz
waveform).
d
Click
Setup > Acquisition...
and make sure the acquisition parameters are set
up as follows:
•
Memory Depth = Automatic
•
Sampling rate = 80 GSa/s (maximum)
•
Sin(x)/x Interpolation = Auto
•
Averaging = Disabled
e
Click
Measure > Add Measurement...
.
f
In the Add Measurement dialog box, select the Vertical RMS measurement
and configure it as follows:
•
Source = Channel 1
•
Measurement Area = Entire Display
•
RMS Type = AC
Enter horizontal
scale