Replacing Assemblies
5
Infiniium V-Series Oscilloscopes Service Guide
93
9
Disconnect each input cable. (When reconnecting, you can determine the
appropriate connections by matching the colored o-ring to the bezel color.
Tighten to 8 in-lbs using a calibrated 5/16” open end torque wrench.)
10
Remove the attenuator power cable from each input channel attenuator as
shown below. Then route the entire cable through the backplane board. The
attenuators are keyed to match the connectors.
Figure 10
Figure 11