Calibration
2
Infiniium V-Series Oscilloscopes Service Guide
21
d
Connect the calibration cable from Cal Out to each channel input as
requested.
e
When instructed, connect the calibration cable from the Cal Out on the front
panel of the oscilloscope to the 1250-1200 SMA(m) to BNC(f) adapter and
then connect the other end of the 1250-1200 adapter to the 8120-1840
BNC cable. Connect the other end of the BNC cable to the Trig In on the rear
of the oscilloscope.
f
A Passed/Failed indication appears for each calibration section. If any
section fails, click the Enable Details box for information on the failures. Also
check the calibration cables.
5
If you are calibrating an MSO model oscilloscope, follow the prompts to
connect the digital channels cable to the accessory I/O connector on the rear
of the oscilloscope and to the calibration connector on the MSO calibrator.
6
When the calibration procedure is complete, click
Close
.
CAUTION
Be sure to orient the digital channels cable with the Keysight part number facing
downward as shown on the rear of the oscilloscope.
CAUTION
Insert the digital channels cable carefully into the MSO calibrator to prevent damage
to the pins. Use the notches on the cable and the MSO calibrator to orient the cable
correctly.
alignment notches
Summary of Contents for Infiniium MSOV 084A
Page 1: ...Keysight Infiniium V Series Oscilloscopes Service Guide...
Page 6: ...6 Infiniium V Series Oscilloscopes Service Guide...
Page 14: ...14 Infiniium V Series Oscilloscopes Service Guide 1 General Information...
Page 22: ...22 Infiniium V Series Oscilloscopes Service Guide 2 Calibration...
Page 52: ...52 Infiniium V Series Oscilloscopes Service Guide 3 Testing Performance...
Page 132: ...132 Infiniium V Series Oscilloscopes Service Guide 5 Replacing Assemblies...
Page 162: ...162 Infiniium V Series Oscilloscopes Service Guide 7 Theory of Operation...
Page 163: ......