Infiniium V-Series Oscilloscopes Service Guide
3
Testing Performance
Connections
Procedure
1
Preset the power meter.
2
Ensure that the power sensor is disconnected from any source and zero the
meter.
3
Connect the power sensor to the power meter's Power Ref connector and
calibrate the meter.
4
Make the connections to oscilloscope channel 1 as shown in the preceding
connection diagram.
5
Set up the power meter to display measurements in units of Watts.
NOTE
•
Connect output 1 of the 11667C splitter to the scope channel N input directly using the
11901A adapter and a connector saver, without any additional cabling or adapters.
•
Connect the power sensor directly to output 2 of the power splitter using the 11901C
adapter without any additional cabling or adapters.
•
Connect the microwave cable directly to output 3 of the power splitter using the 11901C
adapter without any additional cabling or adapters.
•
Minimize the use of other adapters.
•
Ensure the connectors are tightened properly to 8 in-lbs (90 N-cm).
Power sensor
cable
Power sensor
E4413A
Power splitter
11667C
11901C adapter
11901A adapter
Microwave cable
Microwave CW
Generator E8257D
Power meter E4418A or
E4419A
connector saver
Oscilloscope
Summary of Contents for Infiniium MSOV 084A
Page 1: ...Keysight Infiniium V Series Oscilloscopes Service Guide...
Page 6: ...6 Infiniium V Series Oscilloscopes Service Guide...
Page 14: ...14 Infiniium V Series Oscilloscopes Service Guide 1 General Information...
Page 22: ...22 Infiniium V Series Oscilloscopes Service Guide 2 Calibration...
Page 52: ...52 Infiniium V Series Oscilloscopes Service Guide 3 Testing Performance...
Page 132: ...132 Infiniium V Series Oscilloscopes Service Guide 5 Replacing Assemblies...
Page 162: ...162 Infiniium V Series Oscilloscopes Service Guide 7 Theory of Operation...
Page 163: ......