Testing Performance
3
Infiniium V-Series Oscilloscopes Service Guide
33
DC Gain Accuracy Test
CAUTION
Ensure that the input voltage to the oscilloscope never exceeds
±5 V
.
NOTE
The oscilloscope under test must be warmed up (with the oscilloscope application running) for
at least 30 minutes prior to the start of any performance test.
Specifications
DC Gain Accuracy
±2% of full scale at full resolution channel scale (±2.5% for
≤
5 mV/div).
Full scale is defined as 8 vertical divisions. Magnification is used below 7.5 mV/div. Below 7.5
mV/div, full scale is defined as 60 mV. The major scale settings are 1 mV/div, 2 mV/div, 5 mV/div, 10
mV/div, 20 mV/div, 50 mV/div, 100 mV/div, 200 mV/div, 500 mV/div, and 1 V/div.
Equipment Required
Description
Critical specifications
Recommended model/part #
Digital Multimeter
DC voltage measurement accuracy better
than ±0.1% of reading
34411A
Cable Assembly
(2 required)
50
Ω
characteristic impedance, BNC (m)
connectors
8120-1840
Adapter
BNC Tee (m)(f)(f)
1250-0781
Adapter
BNC (f) to dual banana
1251-2277
Connector Saver
3.5 mm (f)-(f), shipped with each V-Series
oscilloscope
2.4 mm (f) to 3.5 mm (f) adapter
54919-68717 (for models
≥
20 GHz)
54916-68716 (for models
≤
16 GHz)
Adapter
(2 required)
BNC (f) to SMA (m) adapter
1250-1200
Summary of Contents for Infiniium MSOV 084A
Page 1: ...Keysight Infiniium V Series Oscilloscopes Service Guide...
Page 6: ...6 Infiniium V Series Oscilloscopes Service Guide...
Page 14: ...14 Infiniium V Series Oscilloscopes Service Guide 1 General Information...
Page 22: ...22 Infiniium V Series Oscilloscopes Service Guide 2 Calibration...
Page 52: ...52 Infiniium V Series Oscilloscopes Service Guide 3 Testing Performance...
Page 132: ...132 Infiniium V Series Oscilloscopes Service Guide 5 Replacing Assemblies...
Page 162: ...162 Infiniium V Series Oscilloscopes Service Guide 7 Theory of Operation...
Page 163: ......