70 Keysight16197A Operation and Service Manual
Service
Operation Check
6-
Short Impedance Check
Upon completion of the open impedance check, carry out a short impedance
check with the fixture set for the SHORT state.
Step 1.
At the DUT measuring position, connect an appropriate shorting bar with the
electrode to secure the fixture for the SHORT state. (See
Figure 3-16 on
page 37
).
Step 2.
Make the settings for the 4291B as follow.
Step 3.
Under this condition, measure Ls at 100MHz and 1GHz, separately.
Step 4.
Confirm that the Ls value is within the typical range shown in
Table 6-7
Short-impedance Measurement Repeatability Check
Repeat short impedance measurement to check for the repeatability of
measurement.
Step 1.
Upon completion of the short impedance check, disconnect the shorting bar
and connect it again with the electrode.
Step 2.
Under the same measuring conditions as given above, measure Ls again at
100MHz and 1GHz, separately.
Step 3.
Confirm that the Ls value is within the typical range shown in
Table 6-7
.
Table 6-6
Settings for the Measuring Instrument (Keysight 4291B)
Measuring Condition
Setting
Measurement parameter
Ls
Start frequency
100MHz
Stop frequency
1GHz
OSC Level
0.5V
Number of points
2
Point averaging factor
16
Points averaging
ON
Table 6-7
SHORT Impedance Check: Typical
Parameter
Frequency
Typical (Absolute value)
Ls
100MHz
2.3nH ± 1nH
a
a. The values given above are common to shorting bars of all sizes.
Ls
1GHz
2.3nH ± 1nH
a
Summary of Contents for 16197A
Page 1: ...Operation and Service Manual Keysight 16197A Bottom Electrode SMD Test Fixture...
Page 22: ...22 Keysight16197A Operation and Service Manual Product Description Product Description 2...
Page 42: ...42 Keysight16197A Operation and Service Manual Operation Connecting and Measuring the DUT 3...
Page 54: ...54 Keysight16197A Operation and Service Manual User Maintenance Assembly Check 4...
Page 72: ...72 Keysight16197A Operation and Service Manual Service Operation Check 6...