Keysight16197A Operation and Service Manual 67
Service
Changing the Orientation of or Replacing the Contact Assembly
Changing the Orientation of or Replacing the Contact Assembly
If the center electrode of the contact assembly is smeared, an unused clean
area of the center electrode can be used simply by changing the orientation of
the contact assembly. This section describes the method for changing the
orientation of or replacing the contact assembly.
To change the contact assembly orientation or replace the contact assembly, it
is necessary to use a 1.5-mm Allen wrench (Keysight part number: 8710-0909)
and a Phillips screwdriver.
Procedure for Changing the Orientation of or Replacing the Contact
Assembly
1.
Take off the guide holder, and remove the device guide and electrode
plate. (See to
Figure 3-4 on page 28
).
2.
Remove the four screws from the upper side of the fixture, and then
remove the contact assembly and flange.
Figure 6-3
Removing the Contact Assembly