Keysight16197A Operation and Service Manual 45
User Maintenance
Cleaning
Cleaning
Smeared electrodes result in poor measurement accuracy and repeatability.
Periodically clean the electrodes to ensure accurate measurements.
Areas Requiring Cleaning
The following areas require cleaning.
— The area of the center electrode that will be in contact with the electrode of
the DUT
— The area of the electrode plate that will be in contact with the electrode of
the DUT
Figure 4-1
Areas Requiring Cleaning
Cleaning Method
Use a cleaning rod (Keysight part number: 5182-7586) to clean the electrodes.
Remove smears on the above-mentioned contact area by wiping with the white
rubber portion of the cleaning rod. Take care not to damage the electrode
parts.