36 Keysight16197A Operation and Service Manual
Operation
Performing Fixture Compensation
3-
Figure 3-14
Raising the Pressure Rod
Step 4.
Using tweezers, place an appropriate shorting bar on the electrode along the
device guide frame.
Figure 3-15
Connecting the Shorting Bar
Each shorting bar is made exclusively for a particular device size. Do not
use a shorting bar of the wrong size.
Handle each shorting bar with tweezers, taking care not to soil it. A soiled
shorting bar may impair measurement accuracy and repeatability.
Take care so that the correct face of the shorting bar is placed in contact
with the electrode.