Keysight 16197A Bottom Electrode SMD Test Fixture
Operation and Service Manual
4
User Maintenance
This chapter describes the operations to perform before starting measurement,
and the procedures necessary to ensure accurate measurements.
Description
Necessity of User Maintenance
Through repeated use, the fixture's measuring performance will gradually
deteriorate over time due to smearing on the contact faces from solder
buildup, and to mechanical wear and deformation of the contact faces
themselves. To continually achieve optimal measurement results, it is
important to keep the contact faces in good condition and take appropriate
action before such wear or deformation occurs. Therefore, the fixture should be
properly cared for by implementing each of the maintenance items shown in
Table 4-1
.
Table 4-1
Maintenance Items
Name
Frequency of
implementation
Contents
Target parts
Cleaning
Several times a day
Cleaning the
fixture
Entire fixture
Electrode
deterioration
check
Before using the
fixture for the first time
after delivery/ After
replacement of parts
Acquisition of
reference
values
a
Center electrode,
Electrode plate
Once a day/Before
fixture compensation
Determining the
deviation from
reference values
a
Center electrode,
Electrode plate
Summary of Contents for 16197A
Page 1: ...Operation and Service Manual Keysight 16197A Bottom Electrode SMD Test Fixture...
Page 22: ...22 Keysight16197A Operation and Service Manual Product Description Product Description 2...
Page 42: ...42 Keysight16197A Operation and Service Manual Operation Connecting and Measuring the DUT 3...
Page 54: ...54 Keysight16197A Operation and Service Manual User Maintenance Assembly Check 4...
Page 72: ...72 Keysight16197A Operation and Service Manual Service Operation Check 6...