28 Keysight16197A Operation and Service Manual
Operation
Changing the Orientation of the Device Guide and Electrode Plate
3-
Changing the Orientation of or Replacing the Device Guide and Electrode Plate
Change the orientation of the device guide and electrode plate so that the
selected device guide frame is located at the measuring position. For a 0603
device, replace the attached device guide and electrode plate with ones
supplied in Option 001.
Step 1.
Loosen the knobs on the sliding block, and remove the pressure unit from the
electrode unit, referring to
Figure 1-4 on page 9
Step 2.
Unscrew the screws that fix the guide holder. Then remove the guide holder,
device guide, and electrode plate (
Figure 3-4
)
Figure 3-4
Removing the Device Guide and Electrode Plate
Step 3.
Mount the device guide and electrode plate so that the selected device guide
frame is located at the measuring position. Ensure that the marks on the
device guide are aligned with those on the electrode plate (
Figure 3-5
). For a
0603 DUT, replace the attached device guide and electrode plate with ones
supplied in Option 001 (
Figure 3-6
).
If it is difficult to take out the electrode plate, turn the electrode unit
upside down to remove the plate. Do not use force to take out the
electrode plate, otherwise it may be deformed.
Summary of Contents for 16197A
Page 1: ...Operation and Service Manual Keysight 16197A Bottom Electrode SMD Test Fixture...
Page 22: ...22 Keysight16197A Operation and Service Manual Product Description Product Description 2...
Page 42: ...42 Keysight16197A Operation and Service Manual Operation Connecting and Measuring the DUT 3...
Page 54: ...54 Keysight16197A Operation and Service Manual User Maintenance Assembly Check 4...
Page 72: ...72 Keysight16197A Operation and Service Manual Service Operation Check 6...