32 Keysight16197A Operation and Service Manual
Operation
Working with Device Guide
3-
Figure 3-9
Inside Diameter of Guide Holder & Movable Range of Pressure Rod
Determine the position of the frame to be prepared, keeping in mind the
following requirements:
• The electrode of the DUT must maintain equal contact with both the center electrode
and electrode plate.
• The electrode of the DUT must have a large contact area, and the electrode space
between the center electrode and electrode plate must be as large as possible.
• The pressure rod must be able to press the center of the DUT to stabilize the DUT
during measurement.
Example of Preparing the Device Guide
Preparing the Device Guide for a 2520 SMD
The figure below shown an example of preparation for a device guide for a
2520 SMD. For a 2520 SMD, the measuring position for a 2012 device is used
(electrode space: 1.5mm).