Keysight16197A Operation and Service Manual 57
Specifications and Supplemental Performance Characteristics
Supplemental Performance Characteristics
Supplemental Performance Characteristics
This section provides useful data on the 16197A. These supplemental
performance characteristics should not be considered specifications.
Additional Error
Additional errors are calculated as follows.
|Z| Measurement
Additional error for Impedance Ze [%] is calculated by substituting the values
in the table below into the following equation.
Ze [%] = ± {A + (Zs/Zx + Yo ×Zx) × 100}
where
A [%]
Test Fixture’s Proportional Error [%]
Yo [S]
Test Fixture’s Open Repeatability [S]
Zs [
]
Test Fixture’s Short Repeatability [
]
Zx [
]
Measured Impedance Value of DUT [
]
where f is frequency (GHz)
D Measurement
Additional error for Dissipation Factor De is calculated by using the additional error for
Impedance Ze [%] as follows.
If Dx
0.1:
De = Ze / 100
If 0.1 < Dx
0.5:
De = (Ze / 100) × (1 + Dx)
where Dx is the measured value of D. It is necessary for Ze to be below 10%.
Zs
(30 + 150 × f ) × 10
-3
[W]
Yo
(2 + 30 × f ) × 10
-6
[S]
A
1.2 × f
2
[%]
D is not expressed as a percentage but as an absolute value.
Summary of Contents for 16197A
Page 1: ...Operation and Service Manual Keysight 16197A Bottom Electrode SMD Test Fixture...
Page 22: ...22 Keysight16197A Operation and Service Manual Product Description Product Description 2...
Page 42: ...42 Keysight16197A Operation and Service Manual Operation Connecting and Measuring the DUT 3...
Page 54: ...54 Keysight16197A Operation and Service Manual User Maintenance Assembly Check 4...
Page 72: ...72 Keysight16197A Operation and Service Manual Service Operation Check 6...