Section 2: Performance verification
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
2-52
DMM7510-905-01 Rev. A / April 2015
Configure Digitize DC current for the 10 µA and 100 µA ranges:
1. Set up the Model DMM7510 for digitize DC current and the range being tested. Make sure
relative offset is disabled.
2. Verify the shunt. With a shunt characterization of 200
kΩ parallel 10 MΩ reference, the DMM
typical measurement is 196.07843
kΩ.
Figure 10: Connections for the 10 µA and 100 µA shunt characterization
3. Connect the calibrator, shunt, Model DMM7510, and reference DMM as shown in the following
figure.