Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Section 2: Performance verification
DMM7510-905-01 Rev. A / April 2015
2-55
Verify digitize current 1 mA range
Description
Calibrator setpoint
(A)
Lower limit
Upper limit
UUT Zero
STANDBY
-7.50E-08
7.50E-08
Full scale (+)
0.001
9.99765E-04
1.00024E-03
Half scale (+)
0.0005
4.99845E-04
5.00155E-04
Half scale (-)
-0.0005
-5.00155E-04
-4.99845E-04
Full scale (-)
-0.001
-1.00024E-03
-9.99765E-04
Verify digitize current 10 mA range
Description
Calibrator setpoint
(A)
Lower limit
Upper limit
UUT Zero
STANDBY
-7.50E-07
7.50E-07
Full scale (+)
0.01
9.99765E-03
1.00024E-02
Half scale (+)
0.005
4.99845E-03
5.00155E-03
Half scale (-)
-0.005
-5.00155E-03
-4.99845E-03
Full scale (-)
-0.01
-1.00024E-02
-9.99765E-03
Verify digitize current 100 mA range
Description
Calibrator setpoint
(A)
Lower limit
Upper limit
UUT Zero
STANDBY
-1.00E-05
1.00E-05
Full scale (+)
0.1
9.99450E-02
1.00055E-01
Half scale (+)
0.05
4.99675E-02
5.00325E-02
Half scale (-)
-0.05
-5.00325E-02
-4.99675E-02
Full scale (-)
-0.1
-1.00055E-01
-9.99450E-02
Verify digitize current 1 A range
Description
Calibrator setpoint
(A)
Lower limit
Upper limit
UUT Zero
STANDBY
-1.10E-04
1.10E-04
Full scale (+)
1
9.99390E-01
1.00
Half scale (+)
0.5
4.99640E-01
5.00360E-01
Half scale (-)
-0.5
-5.00360E-01
-4.99640E-01
Full scale (-)
-1
-1.00
-9.99390E-01
Verify digitize current 3 A range
Description
Calibrator setpoint
(A)
Lower limit
Upper limit
UUT Zero
STANDBY
-4.50E-04
4.50E-04
⅔ scale (+)
2
1.00
2.00
Half scale (+)
1.5
1.00
1.00
Half scale (-)
-1.5
-1.00
-1.00
⅔ scale (-)
-2
-2.00
-1.00