Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Section 4: TSP command reference
DMM7510-905-01 Rev. A / April 2015
4-9
•
ACI Adjustment (
cal.adjust.ac()
)
ACI Step 11, 100 µA 1 kHz, ACI 1 mA range zero
ACI Step 12, 1 mA 1 kHz, ACI 1 mA range full-scale and 10 mA range zero
ACI Step 13, 10 mA 1 kHz, ACI 10 mA range full-scale and 100 mA range zero
ACI Step 14, 100 mA 1 kHz, ACI 100 mA range full-scale, 1 A and 3 A range zero
ACI Step 15, 1 A 1 kHz, ACI 1 A range full-scale
•
ACI Step 16, 2 A 1 kHz, ACI 3 A range full-scale
AC-coupling amplitude adjustment
•
ACV Step 17, 1 V 10 Hz, ACV AC coupling amplitude adjustment
Frequency adjustment
ACV Step 18, 1 V 1 kHz, Squarewave, Frequency adjustment
For detail on how to use
cal.adjust.ac()
, refer to:
•
AC voltage adjustment steps 1 to 10
(on page 3-13)
•
AC current adjustment steps 11 to 16
(on page 3-16)
•
AC coupling adjustment step 17
(on page 3-18)
•
AC frequency adjustment step 18
(on page 3-19)
This command generates an error if:
•
Calibration is locked
•
The step is out of sequence
•
The step does not complete successfully
•
The value that is passed is invalid for the step, out of range, or not needed
Also see
(on page 4-12)
(on page 4-16)
(on page 4-19)