Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Section 3: Calibration
DMM7510-905-01 Rev. A / April 2015
3-13
DC voltage adjustment step 5: 10 kΩ
To run DC adjustment step 5:
1. Retain the connections made in
DC voltage adjustment steps 3 and 4: +10 V and -10 V
2. Allow the instruments and cables to settle for 5 minutes.
3. On the calibrator, enable the
OPR
key and the
EX SNS
key.
4. Ensure that the
OPERATE
display and
EX SNS
keys are illuminated.
5. Source
10 kΩ
and allow the calibrator and Model DMM7510 to properly settle.
DC adjust step 5 supports an allow
ed range of source values from 9.0 kΩ to 11.0 kΩ. The Fluke
5720 calibrator supports only approximately 10 kΩ. As a valid alternative, you can use an external
stable and low-drift characterizer reference resistor.
Ensure
cal.adjust.dc(5, 10.0e3)
uses the calibrator actual source or the external stable and
low-drift characterizer reference resistor value.
6. Send the following command.
cal.adjust.dc(5, 9.999938e3) -- 10K Require HI and LO and 4W SHI and SLO Ft
terminals. Enter the actual calibrator source value.
AC voltage adjustment steps 1 to 10
To set up the equipment for AC adjustment steps 1 to 10:
1. Connect a cable between the calibrator and the Model DMM7510 as shown in the figure below.