Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Section 2: Performance verification
DMM7510-905-01 Rev. A / April 2015
2-33
Verify the frequency
Use the following values to verify the performance of the Model DMM7510. Actual values depend on
published specifications (see
Example reading limit calculation
(on page 2-8)).
Description
Frequency (Hz)
Lower limit (Hz)
Upper limit (Hz)
10 Hz @ 5 V
1.00E+01
9.00
1.01
1 kHz @ 5 V
1.00E+03
9.02
1.03
10 kHz @ 5 V
1.00E+04
9.03
1.04
100 kHz @ 5 V
1.00E+05
9.04
1.05
250 kHz @ 5 V
2.50E+05
2.05
2.05
500 kHz @ 5 V
5.00E+05
4.05
5.05
Simulated thermocouple type J temperature verification
1. Connect the Model DMM7510 HI and LO INPUT terminals to the calibrator HI and LO terminals
as shown in the following figure.
Figure 2: DC voltage 100 mV and thermocouple connections