Section 4: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
4-2
DMM7510-905-01 Rev. A / April 2015
acal.lastrun.internaltemp
This attribute returns the internal temperature of the instrument when auto calibration was run.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
temperature
= acal.lastrun.internaltemp
temperature
The internal temperature
Details
The temperature is displayed in Celsius (°C).
The instrument updates the internal temperature value when the instrument refreshes autozero. If
autozero is set to off or if autozero is not available for the selected function (such as capacitance,
continuity, frequency or period), the internal temperature value is not updated.
Example
print(acal.lastrun.internaltemp)
Returns the internal temperature of the instrument when
auto calibration was last run.
Example output:
63.167084
Also see
(on page 4-3)
(on page 4-6)
(on page 4-3)