Section 2: Performance verification
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
2-66
DMM7510-905-01 Rev. A / April 2015
Digitize current 10 A range verification
To verify the 10 A range:
1. Connect Model DMM7510, calibrator, and amplifier as shown in the following figure.
Ensure that the gauge of the cabling is sufficient to handle 10 A.
2. Set the TERMINALS switch to REAR. Ensure that the orange
R
is displayed.
3. Press the
FUNCTION
key, select the
Digitize Functions
tab, and select
Digitize Current
.
4. Press the
HOME
key.
5. Set the Range to
10A
.
6. Press the
MENU
key.
7. Select
Settings
.
8. Set the Sample Rate to
1000
.
9. Set the Aperture to
Auto
or
1 ms
.
10. Set the Count to
100
.