Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Section 3: Calibration
DMM7510-905-01 Rev. A / April 2015
3-35
--AC frequency adjustment step 18
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
eventlog.clear() -- Clear the error event log and Home screen error triangle.
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Ft, AC Step #18, 1V, 1000Hz")
cal.adjust.ac(18, 1000) -- 1V, 1kHz HI and LO Ft terminals. Enter AFG frequency
value.
code,message,severity = eventlog.next() -- Set variables to print the error
messages.
display.changescreen(display.SCREEN_HOME)
delay(2.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Ft, AC Step #18, 1V, 1000Hz , Done")
delay(4.0)
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Calibrate Enabled")
display.settext(display.TEXT2, "Start Next Cal Step")
print('Error Code =', code)
print('Error Message =', message)
print('Error Severity =', severity)
print('done')