Section 4: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
4-12
DMM7510-905-01 Rev. A / April 2015
cal.adjust.dc()
This function starts the specified DC adjustment step.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
cal.adjust.dc(
step
)
cal.adjust.dc(
step
,
value
)
step
The DC adjustment step to start
value
The value for this adjustment step (if applicable)
Details
The descriptions of these steps are:
DC Step 0: Internal amplifier offset adjustment
DC Step 1: 4-wire zero for DCI 1 A to 3
A, DCV, Digitize DCV, 2W Ω, and 4W Ω
DC Step 2: Open Terminals:
•
DCI, zero 10 µA to 100 mA
•
DCV, 100 V and 1000 V full-scale
•
Resistance, 1 Ω to 1 GΩ full-scale
•
Digitize DCI and DCV full-scale
•
Waveform, analog trigger
•
Frequency, threshold level trigger
DC Step 3: 10 V, DCV 10 V range
DC Step 4: -10 V, DCV 10 V range
DC Step 5: 10 kΩ, Resistance, 4W 10 kΩ range
For detail on how to use
cal.adjust.dc()
, refer to
DC voltage adjustment steps 0 to 5
(on page 3-
This command generates an error if:
•
Calibration is locked
•
The step is out of sequence
•
The step does not complete successfully
•
The value that is passed is invalid for the step, out of range, or not needed
Also see
(on page 4-8)
(on page 4-16)
(on page 4-19)