Section 4: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
4-8
DMM7510-905-01 Rev. A / April 2015
Example
acal.schedule(acal.ACTION_RUN, acal.INTERVAL_1DAY, 8)
Sets auto calibration to run every day at 8 am.
Also see
(on page 2-4)
(on page 4-6)
cal.adjust.ac()
This function starts the specified AC adjustment step.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
cal.adjust.ac(
step
)
cal.adjust.ac(
step
,
value
)
step
The AC adjustment step to start
value
The value for this adjustment step (if applicable)
Details
The descriptions of these tests are:
•
ACV adjustment (
cal.adjust.ac()
)
ACV Step 1, 10 mV 1 kHz, ACV 100 mV range zero
ACV Step 2, 100 mV 1 kHz, ACV 100 mV range full-scale and 1 V range zero
ACV Step 3, 100 mV 50 kHz, ACV 100 mV frequency flatness adjust
ACV Step 4, 1 V 1 kHz, ACV 1 V range full-scale and 10 V range zero
ACV Step 5, 1 V 50 kHz, ACV 1 V range frequency flatness adjust
ACV Step 6, 10 V 1 kHz, ACV 10 V range full-scale and 100 V range zero
ACV Step 7, 10 V 50 kHz, ACV 10 V range frequency flatness adjust
ACV Step 8, 100 V 1 kHz, ACV 100 V range full-scale and 700 V range zero
ACV Step 9, 100 V 50 kHz, ACV 100 V range frequency flatness adjust
ACV Step 10, 700 V 1 kHz, ACV 700 V range full-scale