Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
Section 2: Performance verification
DMM7510-905-01 Rev. A / April 2015
2-57
Verify AC current 1 mA range
Description
Verification point Lower limit
Upper limit
@ 40 Hz
0.001
9.989000E-04
1.001100E-03
@ 1 kHz
0.001
9.989000E-04
1.001100E-03
@ 5 kHz
0.001
9.988000E-04
1.001200E-03
Verify AC current 10 mA range
Description
Verification point Lower limit
Upper limit
@ 40 Hz
0.010
9.989000E-03
1.001100E-02
@ 1 kHz
0.010
9.989000E-03
1.001100E-02
@ 5 kHz
0.010
9.988000E-03
1.001200E-02
Verify AC current 100 mA range
Description
Verification point Lower limit
Upper limit
@ 40 Hz
0.1
9.989000E-02
1.001100E-01
@ 1 kHz
0.1
9.989000E-02
1.001100E-01
@ 5 kHz
0.1
9.988000E-02
1.001200E-01
Verify AC current 1 A range
Description
Verification point Lower limit
Upper limit
@ 40 Hz
1.000
9.976000E-01
1.00
@ 1 kHz
1.000
9.976000E-01
1.00
@ 5 kHz
1.000
9.908000E-01
1.00
Verify AC current 3 A range
Description
Verification point Lower limit
Upper limit
2 A @ 40 Hz
2.000
1.00
2.00
2 A @ 1 kHz
2.000
1.00
2.00
2 A @ 5 kHz
2.000
1.00
2.00