Chapter 7. Calibrating and Verifying Sensors
7.4.6 The Graph View (cont.)
On the graph, the temperature appears on the Y axis, while the highlighted X axis shows the
stabilized IRTD temperature. All the sample points that lie within the deviation specified are
displayed in green dots along with the label and the temperature point of the sensor. All the
samples that deviated from the specification are displayed in red dots along with the label of
the sensor, the temperature point of the sensor and the value that indicates how much the
sensor is off.
On the summary tab, the outliers of all the SIM are displayed, along with the SIM number,
sensor label, the temperature point of the sensor and the value of how much the sensor is off.
7.5
Current Calibration and Hardware Connections
Both List and Graph views display essential data:
Bath serial number
Hardware Connection status
Preview stability criteria
Calibration Checklist -- A calibration checklist allows users to keep track of their
particular step in the calibration:
Temperature Reference Set
Sensors Stabilized
IRTD Temperature standard stabilized
Deviation calculated on uncalibrated sensors
Sensors Calibrated
Deviation calculated on calibrated sensors
Saving Calibrated Temperature to files
Calibration point complete
A green check mark appears next to each step in the process when it is completed.
Kaye Validator User’s Manual
83
Summary of Contents for Validator AVS
Page 1: ...Validation Kaye Validator AVS User s Manual Kaye Validator User s Manual...
Page 2: ...Kaye Validator AVS Thermal Process Validation System User s Manual M5100 EN Rev A May 2015...
Page 3: ...No content intended for this page Kaye Validator User s Manual...
Page 133: ...Chapter 10 Generating Reports 126 Kaye Validator User s Manual...
Page 138: ...Chapter 10 Generating Reports Kaye Validator User s Manual 131...
Page 147: ...Chapter 10 Generating Reports 140 Kaye Validator User s Manual...
Page 167: ...No content intended for this page...