Chapter 10. Generating Reports
10.5 The Reports Details Screen (cont.)
Use the up and down arrows to arrange the reports in ascending or descending order. A
Comments column allows users to review the comments entered for a particular setup. At the
right, the Actions column shows three icons:
•
Graph -- navigates to the Graph Analysis page through the
Select Cycles window>>
Create Cycles
page.
•
Reports -- navigates to the Reports page through the
Select Cycles window>> Create
Cycles
page.
If user Selects Summary Analysis in Report Analysis Screen then from Report Detail
Screen it will navigate to Template Selection Screen
If user Selects Performance Analysis Screen it will ask for Option to Select for unit
Temperature
Humidity
Pressure
Contact
Voltage
Current
Note:
For the setup, calibration and calibration verification reports, only the Print icon is
available. Users can save the report as a PDF file or send it directly to a default printer.
Special rules apply for merging study files
The files must also meet certain criteria:
•
The software can merge up to 3 files maximum.
•
The Sampling rate should be same, or the sampling rate of one file is a multiple of second file
(for instance, three minutes and six minutes)
•
Files must overlap for the study duration
•
The printed merged detail report includes data from all three files.
Kaye Validator User’s Manual
117
Summary of Contents for Validator AVS
Page 1: ...Validation Kaye Validator AVS User s Manual Kaye Validator User s Manual...
Page 2: ...Kaye Validator AVS Thermal Process Validation System User s Manual M5100 EN Rev A May 2015...
Page 3: ...No content intended for this page Kaye Validator User s Manual...
Page 133: ...Chapter 10 Generating Reports 126 Kaye Validator User s Manual...
Page 138: ...Chapter 10 Generating Reports Kaye Validator User s Manual 131...
Page 147: ...Chapter 10 Generating Reports 140 Kaye Validator User s Manual...
Page 167: ...No content intended for this page...