Chapter 9. The Qualification Study
9.3 Viewing the Active Qualification Study (cont.)
•
O
pen, under-range, over-range or failed sensors are not allowed to participate in
calculations or events.
•
Each box lists the SIM number, group name, sensor label and associated measurand
(temperature, humidity, pressure, etc.) If there are more sensors or calculations than fit on
the screen, use the scroll bar to display the remaining data.
•
A button for to stop qualification is available at the bottom of the screen.. Data is
refreshed at the rates shown below:
Table 2: Data Refresh Rates
Sampling Rate
Data Refresh Rate
1 sec
3 to 5 secs
2 sec
5 to 7 secs
5 sec
8 to 10 secs
10 sec
13 to 15 secs
15 sec
18 to 20 secs
20sec
23 to 25 secs
30 sec to 12 hours
sampling Time + (3-5) secs
To stop the study, press the
Stop Qualification
button.
100
Kaye Validator User’s Manual