Chapter 10. Generating Reports
• Lethality
• Sat P
• Sat T
• MKT calculations
Text boxes enable you to select the minimum and maximum values of the Y axis for the
selected calculation graph. (These boxes accept only numerics, along
with the minus sign, decimal point and comma.)
Above the Groups/Sensors pane, three tabs allow you to select the Lethality, Saturation
Pressure and Saturation Temperature calculations.
• The Lethality calculation is available only if temperature sensors are present in the study.
The left pane has the lethality graph for the entire study, while
the right pane enables you to change the D, Z and Base temperature values.
• The Saturation Pressure calculation is available only if the study includes at least one
temperature sensor along with pressure sensors.
• The Saturation Temperature calculation is available only if the study includes at least one
pressure sensor along with temperature sensors.
• No calculations are available for a contact sensor group.
Based on the group selection, the following calculation graphs are generated:
• Temperature Sensors
• Temperature Sensors Calculations
• RH (Humidity) Sensors
• RH Sensors Calculations
• Pressure Sensors
134
Kaye Validator User’s Manual
Summary of Contents for Validator AVS
Page 1: ...Validation Kaye Validator AVS User s Manual Kaye Validator User s Manual...
Page 2: ...Kaye Validator AVS Thermal Process Validation System User s Manual M5100 EN Rev A May 2015...
Page 3: ...No content intended for this page Kaye Validator User s Manual...
Page 133: ...Chapter 10 Generating Reports 126 Kaye Validator User s Manual...
Page 138: ...Chapter 10 Generating Reports Kaye Validator User s Manual 131...
Page 147: ...Chapter 10 Generating Reports 140 Kaye Validator User s Manual...
Page 167: ...No content intended for this page...