Chapter 7. Calibrating and Verifying Sensors
7.4.5 Data Logging Complete at Setpoint
When the Kaye Validator finishes logging the corrected results, it automatically sets the Kaye
temperature reference to the high setpoint and repeats the calibration process. When the
calibration process is complete for the high setpoint, the Kaye Validator automatically sets the
Kaye temperature reference to the check setpoint. The only difference between the low and
high setpoint calibration and the check setpoint is that calibration offsets are not changed
during check setpoint.
7.4.6 The Graph View
In the calibration study Graph View, the header consists of icons representing four SIMs,
along with a summary. A drop-down list below these icons displays the setpoint
Figure 41: Calibration Study -- Graph View
82
Kaye Validator User’s Manual