Chapter 7. Calibrating and Verifying Sensors
7.3 Selecting Sensors (cont.)
After you have selected and confirmed your sensors, press
Start Calibration
to start a new
calibration, or
Verify Calibration
to perform a verification. (Press
Cancel
to discard the
sensor selections and return to a fresh screen. Press the X in the upper right to exit the option.)
Note:
Once you begin a new calibration, any offsets from previous calibrations will be lost.
7.4 Calibrating or Verifying Sensors
The Kaye Validator automatically calibrates (or verifies) sensors at the setpoints defined in
your setup.
To calibrate sensors, from the Select Sensors screen, press
Calibrate Sensors
. To verify
sensors, from the Select Sensors screen, press
Verify Calibration
.
Figure 40: Calibration Study Screen
Kaye Validator User’s Manual
79