Chapter 6. Defining Study Setups
Chapter 6. Defining Study Setups
Before you can run a qualification study, you must use the Kaye Validator software to create
or modify a setup. A setup defines everything required to calibrate sensors and run a
qualification study.
Note:
To create or modify a setup, you must have permissions established by your System
Administrator in the Admin menu (User Management tab).
You access setups from the Asset Details screen, through the Setup pane. When you create a
setup, you:
•
Define the sensors you are going to use in the study
•
Assign sensors to groups and specify group calculations
•
Define group events to be monitored during the qualification study
•
Enter group-specific information to be used in report headers
•
Specify calibration setpoints
•
Specify temperature stability and deviation criteria for sensor calibration
•
Specify start and stop conditions for the qualification cycle and the exposure cycle
•
Define the output relay
After you have created and saved your setup, you can calibrate sensors and run a
qualification study.
Kaye Validator User’s Manual
59
Summary of Contents for Validator AVS
Page 1: ...Validation Kaye Validator AVS User s Manual Kaye Validator User s Manual...
Page 2: ...Kaye Validator AVS Thermal Process Validation System User s Manual M5100 EN Rev A May 2015...
Page 3: ...No content intended for this page Kaye Validator User s Manual...
Page 133: ...Chapter 10 Generating Reports 126 Kaye Validator User s Manual...
Page 138: ...Chapter 10 Generating Reports Kaye Validator User s Manual 131...
Page 147: ...Chapter 10 Generating Reports 140 Kaye Validator User s Manual...
Page 167: ...No content intended for this page...